| =Introduction= | | =Introduction= |
− |
| |
− | Wireless sensor "Callibri" is a multifunction device that can be used for:
| |
− |
| |
− | :1. registration of various biopotentials (electromyograms, electrocardiograms, cardiointervalograms);
| |
− |
| |
− | :2. for recording the angular velocity and acceleration (MEMS);
| |
− |
| |
− | :3. the formation of electrostimulating effects and correction of violations of the human muscular system.
| |
− |
| |
− | The principle of wireless communication is provided using the Bluetooth LE radio channel between the sensor and the device (mobile phone, etc.)
| |
− | The advantage of wireless registration is to minimize motor artifacts and, as a consequence, to ensure the free position of the subject's body, including the ability to move.
| |
− | Based on the sensors "Callibri" with the use of SDK, you can create:
| |
− |
| |
− | :• neurointerfaces based on Biological feedback;
| |
− |
| |
− | :• a system for assessing the quality of training in fitness and sports;
| |
− |
| |
− | :• applications based on the registration of EMG, EEG, ECG (no more than 4 channels).
| |
− |
| |
− | To start develop your apps using Neurotech Callibri SDK read <u>[[Developer guide]]</u>, <u>[[Developer guide V2]]</u>
| |
− |
| |
− | [[Product "Callibri"]]
| |
− |
| |
− | ==Appearance of the sensor "Callibri"==
| |
− |
| |
− |
| |
− | '''1. Power button''' [[File:4.png|thumb|right|sensor "Callibri"]]
| |
− |
| |
− | '''2. Red indicator'''
| |
− |
| |
− | '''3. Green indicator'''
| |
− |
| |
− | '''4. Cutout in the sensor for tape attachment'''
| |
− |
| |
− | '''5. Contact areas'''
| |
− |
| |
− | '''6. USB connector for the charger'''
| |
− |
| |
− | ==Start and finish of work==
| |
− |
| |
− | Press the power button for turning on/off the sensor (1).
| |
− |
| |
− | ==Charging==
| |
− |
| |
− | For charge the sensor align the sensor's USB connector (6) with the USB cable of the charger. Connect to power supply.
| |
− |
| |
− | [[File:5.png]]
| |
− |
| |
− | <big>'''WARNING! DO NOT USE A SENSOR DURING THE CHARGE'''</big>
| |
− |
| |
− | [[File:1.png]] Use only chargers with appropriate specifications. Use of inappropriate chargers may result to the battery explosion or sensor damage.
| |
− |
| |
− | [[File:2.png]] Charge the battery before using the "Callibri" registration module for the first time!
| |
− |
| |
− | [[File:2.png]] Charging of the registration module is ONLY when it is turned off! Otherwise, there is a probability that the "Callibri" sensor will fail.
| |
− | Do not use a personal computer to charge the "Callibri" sensor!
| |
− |
| |
− | [[File:2.png]] Can not keep the registration module in a discharged state!
| |
− |
| |
− | [[File:3.png]] The duration of the full charge of the battery is no more than 4 hours.
| |
− |
| |
− | ==Indication of sensor "Callibri"==
| |
− |
| |
− | On body of the sensor "Callibri" there are two indicators of green and red colors. The color indication depends on the operating mode of the sensor.
| |
− |
| |
− |
| |
− | [[File:a.png]] - LEDs do not light up - the registration module "Callibri" is disabled;
| |
− |
| |
− | [[File:b.png]] - The LED on the left is green - the "Callibri" registration module is on;
| |
− |
| |
− | [[File:c.png]] - The LED left blinks in green - the process of charging the battery of the "Callibri" registration module is in progress;
| |
− |
| |
− | [[File:d.png]] - The right LED is lit in red - the battery charge of the "Callibri" registration module is less than 5%;
| |
− |
| |
− | [[File:e.png]] - The right LED flashes red - the battery charge of the "Callibri" registration module is less than 1%;
| |
− |
| |
− | [[File:f.png]] - LED left and right alternately flash green and red colors - the registration module "Callibri" is fully charged.
| |
− |
| |
− |
| |
− | ==Product using==
| |
− |
| |
− | Wireless sensor "Callibri" can be used to:
| |
− |
| |
− | :1) diagnosis;
| |
− |
| |
− | :2) electrostimulation / correction.
| |
− |
| |
− | <big>'''WARNING! WHEN A SENSOR IS USING IN ELECTRIC MODULATION / STIMULATION, FOLLOW THE SAFETY REQUIREMENTS WHEN USING THE DEVICE!'''</big>
| |
− |
| |
− | For the wireless sensor "Callibri in the mode of electrostimulation / correction:
| |
− |
| |
− | :1. Contraindications:
| |
− |
| |
− | ::1.1 individual intolerance to electric current;
| |
− |
| |
− | ::1.2 damage to the integrity of the skin (wounds, abrasions, ulcers);
| |
− |
| |
− | ::1.3 Built-in pacemaker, neurostimulator or any other electronic device;
| |
− |
| |
− | ::1.4 with caution in the presence of the patient epileptic seizures. Only with the permission of the attending physician!
| |
− |
| |
− | ::1.5 heart disease. Only after consulting a doctor;
| |
− |
| |
− | ::1.6 serious violations of the arterial blood circulation in the lower extremities.
| |
− |
| |
− | :2. Not recommended:
| |
− |
| |
− | ::2.1 for prolonged electrostimulation (more than 30 minutes);
| |
− |
| |
− | ::2.2 Use the appliance by pregnant women.
| |
− |
| |
− | :3. Prohibited:
| |
− |
| |
− | ::3.1 use the product near the head. There is no evidence of the effects of brain stimulation;
| |
− |
| |
− | ::3.2 use the product to stimulate the neck. Possible consequences in the form of severe muscle spasms and closure of the airways;
| |
− |
| |
− | ::3.3 perform stimulation on the chest. Possible consequences in the form of a violation of the rhythm of the heart.
| |
− |
| |
− | ==Preparation of a wireless sensor "Callibri" for working==
| |
− |
| |
− | :'''For work with the "Callibri" complex you should:'''
| |
− |
| |
− | :1. Enable the Bluetooth module on the mobile device.
| |
− |
| |
− | :2. Turn on the registration module "Callibri" by touching and holding the button for two seconds.
| |
− |
| |
− | :3. If you want use the diagnostic mode.
| |
− |
| |
− | ::3.1 install disposable electrodes on the "Callibri" registration module;
| |
− |
| |
− | ::3.2 remove the protective layer from the electrodes;
| |
− |
| |
− | ::3.3 install the electrodes:
| |
− |
| |
− | :::• abdominal muscles (for EMG);
| |
− |
| |
− | :::• on the non-haired part of the head (frontal bone) (for EEG);
| |
− |
| |
− | :::• the area of the projection of the heart on the patient's chest (for HRV).
| |
− |
| |
− | ::3.4 after finishing pick up of the signals, remove the electrodes from the body;
| |
− |
| |
− | ::3.5 turn off the registration module.
| |
− |
| |
− | :4. If you want use the electro-stimulation / correction mode:
| |
− |
| |
− | ::4.1 put the cuff on the limb (upper / lower);
| |
− |
| |
− | ::4.2 place the registration module "Callibri" in the pocket on the cuff;
| |
− |
| |
− | ::4.3 connect the cable to the registration module "Callibri" on the connectors (red and blue), install disposable electrodes;
| |
− |
| |
− | ::4.4 remove the protective layer from the electrodes;
| |
− |
| |
− | ::4.5 place the electrodes on the abdomen of the necessary muscle;
| |
− |
| |
− | ::4.6 perform stimulation / correction;
| |
− |
| |
− | ::4.7 remove electrodes from the body;
| |
− |
| |
− | ::4.8 disconnect the cable from the electrodes and the registration module;
| |
− |
| |
− | ::4.9 remove the registration module from the pocket on the cuff and turn it off;
| |
− |
| |
− | ::4.10 remove the cuff.
| |
− |
| |
− | :5. Utilize the disposable electrodes by a standard method.
| |
− |
| |
− | :6. Disinfect the registration module with a tissue moistened with a 3% hydrogen peroxide solution, with the addition of a synthetic detergent.
| |